Operator: Chris Lilley (svgeesus) Date: 10 Nov 2018 DUT: US Cal Club 2 PX reference Type 1b: Direct measurement, calibrated equipment (in 1 year spec or better) DMM: Keysight 34465A, instrument ID 869657 Cal date: 28 Nov 2017 Cal type: Accredited 17025 with uncertainties Cal temp: 23.0C Cal humidity: 23.0% RH Cal procedure: 34460-90901 Truevolt DMM service manual, July 2016 edition 3 Cal inst: Fluke 5725A amplifier, ID 06-0090, NIST 12678290, Cal date 03 Aug 2017 Cal inst: Fluke 5700A/EP03 calibrator, ID 21-0140, NIST 11901768, Cal date 01 Dec 2016 Cal inst: Agilent/HP 33250A function generator, ID 407410002, NIST 12901089, Cal date 13 Oct 2017 Cal front panel zero offset, short, 10V range: as left 0.00000, min -0.00002V max 0.00002V, uncertainty 5.8uV Cal front panel +10V, 10V range: as left 9.99999V, min 9.99966V max 10.00034V, uncertainty 35uV Measurement details: front panel, 100PLC, HiZ PX2 mean: 7.0456366V PX2 lowest: 7.04563V PX2 highest: 7.04564V PX2 samples: 562 PX2 sd: 1uV PX2 corrected: 7.045643V ambient: 22.0C 33% RH (arduino, free air) 22.5 (k-type, above ref) power: 11.94688V from supplied voltage regulator, fed with 15.0V from Tenma 72-8335A linear bench PSU