B2980A 2.3.2011.5670 RELEASE NOTE - FIXED DEFECTS #535954 Need to show message that List Data is not kept, when # of data is over 2500. #579971 No error message when storing over 2.5k points list sweep data. #580514 Memory leak on socket/5025 connections. B2980A 2.3.1924.2760 RELEASE NOTE - FIXED DEFECTS #572658 High voltage self-test improvements. B2980A 2.3.1916.4990 RELEASE NOTE - FIXED DEFECTS #447095 Unexpected exception on power-up. B2980A 2.3.1902.5305 RELEASE NOTE - FIXED DEFECTS #48623 Lost model/serial number in the non-volatile memory. B2980A 2.2.1834.5220 RELEASE NOTE - FIXED DEFECTS #447095 Unexpected exception on power-up. #547092 Wants to reset automatically when Easy File Access status was changed by 'Factory Reset'. #547169 Slow UI response on "Load List Sweep Data" dialog when USB stick contains large waveform data. #557047 Unexpected voltage output on sequential measurement. B2980A 2.2.1745.7190 RELEASE NOTE - FIXED DEFECTS #501340 Cannot restore output low terminal configuration by *RCL. #501500 [SCPI] :TRAC:STAT:FORM causes an error. B2980A 2.1.1645.5880 RELEASE NOTE - FIXED DEFECTS #160933 Unexpected SCPI error message when invalid channel suffix was specified. #447095 Unexpected exception on power-up. #480365 Unexpected exception occurs during infinite sampling measurement. #481907 "Auto Clear" setting in Composite Limit Test Setup Dialog doesn't work. #482463 EOT in Limit test function doesn't work when auto clear function is disabled. #484099 Unexpected exception occurs when Limit Test is triggered with END update. #484100 Up/Low Pattern is reset if up/low limit is updated in Limit Test Setup dialog. B2980A 2.0.1608.8800 RELEASE NOTE - NEW FEATURES #318808 Real-time noise monitor. #372695 Support absolute value of resistance measurement. #372700 Return actual voltage value when performing voltage measurement with N1413A/N1414A adapters. #389137 Wants to show measurement result in exponential format. #389140 Needs sequential measurement controls for resistance measurement standards. #390996 Wants more precise digits for resistance measurement result in engineering mode. #390998 Wants to set longer interval on logger application. - FIXED DEFECTS #330077 Cannot apply sweep voltage without trigger delay. #447095 Unexpected exception on power-up. #450043 B2985A sometimes crashed. #457216 Cannot set data output format (data type) on the front panel. B2980A 1.1.1527.1533 RELEASE NOTE - FIXED DEFECTS #372097 Device state is not recalled correctly at power turn-on. #374402 PLC measurement aperture is not updated when PLC Auto Detect is selected in auto-triggering mode. #387355 [Easy File Access] OK button may not be shown on configuration changes. #389135 Unexpected ohm range changes by navigation keys when ohm ranging is disabled by voltage source control mode. #403781 [Easy File Access] Incorrect PNG filename is shown to the web browsers. #403784 [Easy File Access] Empty folder may be shown to the explorer. B2980A 1.1.1508.3400 RELEASE NOTE - FIXED DEFECTS #330855 Wants to recover trigger mode on GTL. #340615 Needs to add wait time between VF DAC change and output state change. #343628 Cannot check the revision of VF FPGA. #344934 Memory leak on socket connection. #346500 Unexpected discharge state. #346501 Unexpected squeare waveform output if end-time is 0. #346502 Unexpected message is shown when Y axis data element is changed. #351053 Unexpected recall result on Graph view. #352209 Unexpected VF emergency. #355704 Needs discharge operation after input relay is turned off. #355717 Needs discharge operation on charge measure range changes. #355719 Unnecessary wait time on auto ranging. #355723 Need to shorten wait time on auto ranging. #356145 Java code signing #359735 Need to wait before single measurement after input relay is turned on. #360307 Wants to reset display properties. #360545 Output switch LED is dark. #361541 Temperature sensor on VF board cannot measure negative value. #362084 Wants self-test fail log #362135 Unexpected resistance measurement result by null function. #362524 Wants CRC check function between main board and VF board. #364750 Power-on self-test fails on low temperature. #365444 Assist keys don't work for trigger count. #367391 NaN data is shown as temperature data if thermocouple was not calibrated. #372096 Variance ratio value is not displayed correctly. B2980A 1.0.1430.6131 RELEASE NOTE - INITIAL RELEASE