Wavetek 4953 10A AC/DC precision shunt review

Contents

Intro

This is stub for Wavetek 4953 Shunt. This shunt is designed and sold in bundle with Wavetek 4950 Multi-function Transfer Calibrator.

xDevs.com Wavetek 4953 Fluke A40B-10A Units
Nominal shunt current 0 to 11 0 – 10 A
Nominal DC resistance 0.01 0.08 Ω
Input / output connector 5-way binding posts N female connector
Frequency range DC, 10 – 20k DC, 10 – 100k Hz
Short-term stability, ±1 °C ± 20 ± 26 ppm of FS
Temperature coefficient 10 10 ± ppm/°C
AC Flatness Low band 200 ppm, 10 Hz – 1 kHz 26 to 37 ppm, 10 Hz – 1 kHz
AC Flatness Hi band 300 ppm 5kHz, 600 ppm 10 kHz, 1000 ppm, 20 kHz 60 ppm 10kHz, 92 ppm 100 kHz
Temperature coefficient, AC 40 ppm Low band, up to 120 ppm Hi band 4.5 ppm

Table 1: Specification of the Wavetek 4953

Disclaimer

Redistribution and use of this article or any images or files referenced in it, in source and binary forms, with or without modification, are permitted provided that the following conditions are met:

  • Redistributions of article must retain the above copyright notice, this list of conditions, link to this page (https://xdevs.com/article/w4953/) and the following disclaimer.
  • Redistributions of files in binary form must reproduce the above copyright notice, this list of conditions, link to this page (https://xdevs.com/article/w4953/), and the following disclaimer in the documentation and/or other materials provided with the distribution, for example Readme file.

All information posted here is hosted just for education purposes and provided AS IS. In no event shall the author, xDevs.com site, or any other 3rd party be liable for any special, direct, indirect, or consequential damages or any damages whatsoever resulting from loss of use, data or profits, whether in an action of contract, negligence or other tortuous action, arising out of or in connection with the use or performance of information published here.

If you willing to contribute or have interesting documentation to share regarding pressure measurements or metrology and electronics in general, you can do so by following these simple instructions.

Exterior and connections, S/N 36149


Image 1-2. Wavetek 4953 AC/DC 10A shunt facepanel


Image 3. Original labels on the bottom of the shunt


Image 1-2. Owner’s calibration results

Teardown of unit S/N 324XX

This unit was evaluated some years back and found to be functional.

Also rare chance presented to look inside of the shunt for internal construction review.

Pretty large shunt element was placed between two dielectric plates. Wiring to the current input ports is done with heavy braid to lower inductance and AC errors from the connection.

Output is routed to output therminals with two short wires soldered to tab rings. Joints at the shunt secured with brown epoxy for mechanical stability.

Shunt element marked as 408941-1 with date code 14th week 1996.

Calibration test, S/N 36272

To perform actual performance test on this model shunt we used yet another sample, S/N 36272 from a friendly lab here in USA.

Top looks exactly like the other two, no surprises here.

To maintain any history and repeatability of results no disassembly or any intrusive operations were done to this shunt. Connectors were cleaned with 99% isopropyl alcohol and light brushing with paper prior to measurements.

Model and serial Calibration date Test current Assigned value Uncertainty Measured deviation Notes
W4953 36272 27.OCT.2023 by xDevs 0.245 A DC 0.010008608 Ω ± 1.2 µΩ/Ω REFERENCE 10 hours, 6010B DCC vs MPx-0R1
W4953 36272 29.OCT.2023 by xDevs 0.245 A DC 0.010008573 Ω ± 2.5 µΩ/Ω -3.5 µΩ/Ω 10 hours, Fluke 5720A + HPAK 34420A
W4953 36272 30.OCT.2023 by xDevs 0.500 A DC 0.010008606 Ω ± 3 µΩ/Ω -0.2 µΩ/Ω 24 hours, Fluke 5720A + HPAK 34420A
W4953 36272 30.OCT.2023 by xDevs 1.0 A DC 0.010008620 Ω ± 4 µΩ/Ω +1.2 µΩ/Ω 24 hours, Fluke 5720A + HPAK 34420A
W4953 36272 1.NOV.2023 by xDevs 2.0 A DC 0.010008697 Ω ± 6 µΩ/Ω +8.9 µΩ/Ω 24 hours, Fluke 5720A + HPAK 34420A
W4953 36272 3.NOV.2023 by xDevs 4.0 A DC 0.010008980 Ω ± 6 µΩ/Ω +37 µΩ/Ω 24 hours, Fluke 5720A + HPAK 34420A
W4953 36272 5.NOV.2023 by xDevs 6.0 A DC 0.010009052 Ω ± 8 µΩ/Ω +44 µΩ/Ω 24 hours, Fluke 5720A + HPAK 34420A
W4953 36272 7.NOV.2023 by xDevs 8.0 A DC 0.010009117 Ω ± 8 µΩ/Ω +51 µΩ/Ω 24 hours, Fluke 5720A + HPAK 34420A
W4953 36272 8.NOV.2023 by xDevs 10.0 A DC 0.010009226 Ω ± 8 µΩ/Ω +62 µΩ/Ω 24 hours, Fluke 5720A + HPAK 34420A
W4953 36272 8.NOV.2023 by xDevs 2.0 A DC/AC 1kHz 0.010011435 Ω ± 60 µΩ/Ω +274 µΩ/Ω AC/DC 6 hours, Fluke 5720A + Fluke 5790A, INPUT2 HiZ
W4953 36272 8.NOV.2023 by xDevs 4.0 A DC/AC 1kHz 0.010012433 Ω ± 54 µΩ/Ω +345 µΩ/Ω AC/DC 4 hours, Fluke 5720A + Fluke 5790A, INPUT2 HiZ
W4953 36272 8.NOV.2023 by xDevs 6.0 A DC/AC 1kHz 0.010012512 Ω ± 52 µΩ/Ω +346 µΩ/Ω AC/DC 4 hours, Fluke 5720A + Fluke 5790A, INPUT2 HiZ
W4953 36272 10.NOV.2023 by xDevs 10.0 A DC/AC 10 Hz 0.010009381 Ω ± 48 µΩ/Ω +16 µΩ/Ω AC/DC 8 hours, Fluke 5720A + Fluke 5790A, INPUT2 HiZ
W4953 36272 8.NOV.2023 by xDevs 10.0 A DC/AC 40 Hz 0.010009422 Ω ± ? µΩ/Ω +20 µΩ/Ω AC/DC 8 hours, Fluke 5720A + Fluke 5790A, INPUT2 HiZ
W4953 36272 9.NOV.2023 by xDevs 10.0 A DC/AC 60 Hz 0.010009456 Ω ± ? µΩ/Ω +23 µΩ/Ω AC/DC 8 hours, Fluke 5720A + Fluke 5790A, INPUT2 HiZ
W4953 36272 10.NOV.2023 by xDevs 10.0 A DC/AC 100 Hz 0.010009470 Ω ± ? µΩ/Ω +24 µΩ/Ω AC/DC 8 hours, Fluke 5720A + Fluke 5790A, INPUT2 HiZ
W4953 36272 9.NOV.2023 by xDevs 10.0 A DC/AC 300 Hz 0.010010176 Ω ± ? µΩ/Ω +94 µΩ/Ω AC/DC 8 hours, Fluke 5720A + Fluke 5790A, INPUT2 HiZ
W4953 36272 8.NOV.2023 by xDevs 10.0 A DC/AC 1 kHz 0.010012817 Ω ± 45 µΩ/Ω +359 µΩ/Ω AC/DC 8 hours, Fluke 5720A + Fluke 5790A, INPUT2 HiZ
W4953 36272 10.NOV.2023 by xDevs 10.0 A DC/AC 5 kHz 0.010022438 Ω ± ? µΩ/Ω +1320 µΩ/Ω AC/DC 10 hours, Fluke 5720A + Fluke 5790A, INPUT2 HiZ
W4953 36272 11.NOV.2023 by xDevs 10.0 A DC/AC 10 kHz 0.010036484 Ω ± ? µΩ/Ω +2724 µΩ/Ω AC/DC 13 hours, Fluke 5720A + Fluke 5790A, INPUT2 HiZ

Temperature stability

Frequency flatness comparisons

To compare AC/DC transfer performance set of additional shunts was used:

  • Fluke A40B 20A shunt
  • Fluke A40A 20A shunt

Summary & Conclusion

TBD

Author: Ilya Tsemenko
Created: Feb. 13, 2020, 6:03 a.m.
Modified: Nov. 11, 2023, 6:12 p.m.