Volt transfer experiment, MM -> xDevs.com, by using A9 PCBA
732B Reference level: 9.99995638 (9.99996426) VDC
CSV-data from HP 3458A to 732B Transfer, ACALFluke 732B specification
1 Year : ± 2.0 ppm
HPAK 3458A-002 specification (Calibrated 6/2015, 7 months from now):
Accuracy 10V: 4 ppm of reading + 0.05 ppm of range, 24 hour, ±1° C of last ACAL; TAMB = Tcal ±5°C, MATH NULL and fixed range
Transfer accuracy 10V: 0.05 ppm of reading + 0.05 ppm of range, 10 minute window, ±0.5°C TAMB