Test with nVmeter diffs, 2182 CH1 in opposition to 732B-CELL2

Datalog run GPIB code running on Pi3 + E5810A with DP320B scanner

All values reported versus SI readings
732B CELL1 output = reference 10.000002541 VDC from JJ transfer, December 2020
732B CELL2 output = reference 9.999925027 VDC from JJ transfer, December 2020 [Key standard]
732B CELL3 output = reference 9.999928326 VDC from JJ transfer, December 2020
732B CELL4 output = reference 9.999928414 VDC from JJ transfer, December 2020
7004 CELL1 output = reference 9.999897776 VDC from JJ transfer, December 2020
7004 CELL2 output = reference 9.999931992 VDC from JJ transfer, December 2020
7004 CELL3 output = reference 10.011973010 VDC from JJ transfer, December 2020
7004 AVG output = reference 10.003932759 VDC from JJ transfer, December 2020
Result CSV-data

Workflow:

* RPI accesses DMMs via GPIB and reads data.
* Data is formatted by python script into DSV string and written/append to file on FTP
* DSV-file is visible publicly on https://xdevs.com/datashort/
* Page http://xdevs.com/datalog_mm/ runs D3.js javascript library to read DSV file test.log to input data
* D3.js plots SVG graph online :)