Wirewound Vishay 20K black test ambient temp

Test 20K Vishay wirewound resistor in ambient. 3458A@NPLC100 OCOMP ON with 5s delay

First test

TCBOX is 16 ppm/K


Result CSV-data from HP 3458A

Workflow:

* RPI accesses DMMs via GPIB and reads data.
* Data is formatted by python script into DSV string and written/append to file on FTP
* DSV-file is visible publicly on https://xdevs.com/datashort/
* Page http://xdevs.com/datalog_mm/ runs D3.js javascript library to read DSV file test.log to input data
* D3.js plots SVG graph online :)

Setup photo: